Measuring range: 0 ~ 10KPa
Resolution: 1pa
Measuring accuracy: <1%
Test time: 15 seconds on average
Test mouth diameter: 7.7mm
Insertion depth: 9mm
Power requirements: AC 100-240V,50/60Hz
PC communication: USB port connection
Dimensions: L520 x W280 x H300mm
Weight: 21kg
Professional Technology
Use high precision, large range of imports of electronic flow sensors and pressure sensors, test
cigarette draw resistance performance.
One key test, the sample is installed, simply press a key, you can quickly test the instrument with a
USB interface, can use the included software, statistics, analysis of test results, print the standard test report can also save the test report to EXCEL format.
Test standard
ISO6565,GB/T 22838.5,GB/T 18767
Test principle
Inhalation of a fixed flow simulation on the cigarette smoking act 17.5ml / s, the flow of air, this time will be the suction side of the cigarette to produce a certain vacuum pressure, this pressure is the
inhalation resistance, referred to as the draw resistance, measure the pressure. Can measure the
draw resistance.
Application
Cigarette factory production quality control, cigarette testing, the field of tobacco research.
Configuration
Standard configuration: host, a power line.
Optional accessories: 1.LABTest analysis software. Diameter test port.